Received: 2 March 2019
Revised: 5 May 2019
Accepted: 6 May 2019
DOI: 10.1002/sscp.201900021
RESEARCH ARTICLE
Comparative study of different chemistries and particle
properties, high-performance liquid chromatography stationary
phases in separation of escitalopram oxalate and its impurities in
different pharmaceutical dosage forms
Wafa. F. S. Badulla1,2
Nafiz Öncü Can1
Zeki Atkosar1
Göksel Arli1
Hassan Y. Aboul-Enein3
1 Department
of Analytical Chemistry,
Faculty of Pharmacy, Anadolu University
26470, Eskisehir, Turkey
2 Department
of Analytical Chemistry, Faculty
of Pharmacy, Aden University, Aden, Yemen
3 Pharmaceutical
and Medicinal Chemistry
Department, Pharmaceutical and Drug
Industries Research Division, National
Research Centre, Cairo, 12622, Egypt
Correspondence
Professor Hassan Y. Aboul-Enein, Pharmaceutical and Medicinal Chemistry Department,
Pharmaceutical and Drug Industries Research
Division, National Research Centre, Dokki,
Cairo 12622, Egypt.
Email: haboulenein@yahoo.com
The liquid chromatographic separation performance of 12 different columns was evaluated for the estimation of the escitalopram oxalate and six of its impurities in different
pharmaceutical formulations. All columns exhibited a reasonable separation power.
Six columns with three different chemistries and two different particle morphology
(i.e. core-shell and fully porous particle) were nominated for further evaluation. The
objectives of the current study were to develop a simple method for detection of escitalopram oxalate and its impurities and to compare the separation power of the stationary phase of different chemistry and particle properties. The separation was carried out under isocratic elution by using acetonitrile/methanol/water/phosphate buffer
solution (pH = 3.5, 50 mM) (25:5:20:50, v/v/v/v) mixture as mobile phase, with a flow
rate of 1.2 mL min−1 . The core-shell phenyl-hexyl column showed the best separation
performance over the traditional octadecyl and pentafluorophenyl phases. The developed method was applied for detection of escitalopram oxalate and its impurities in
valid and expired dosage forms.
KEYWORDS
core shell columns, escitalopram oxalate, impurities, pentafluorophenyl columns, porous particles
1
I N T RO D U C T I O N
Escitalopram oxalate (ESC-OX) was approved by the Federal Drug Agency (FDA) as a selective serotonin reuptake
inhibitor in 2002. The S enantiomer of the racemic mixture
of the antidepressant drug citalopram which possessed more
Article Related Abbreviations: ANOVA, analysis of variance; CIT A,
citalopram A; CIT B, citalopram B; CIT C, citalopram C; CIT D,
citalopram D; CIT E, citalopram E; DAD, photo diode array detector;
ESC-OX, escitalopram oxalate; FDA, Federal Drug Agency; K’, capacity
factor; MeOH, methanol; octadecyl, C18; OX-A, oxalic acid; PFP,
pentafluorophenyl; USP, United State Pharmacopeia; α, selectivity of
columns.
Sep Sci plus 2019;1–16.
potency and low side effect compared to the R enantiomer [1].
Control of the pharmaceutical impurities became a critical
demand for the determination of drug safety and efficacy [2].
The analytical methods relating to the estimation of ESCOX impurities are very limited including HPLC, LC-MS/MS
and capillary electrophoresis techniques [3–11]. Most of these
methods were based on quantification and qualification of the
R-enantiomer or other impurities that were not listed in the
official USP monograph of the ESC-OX. The latest version of
United State Pharmacopeia (USP 2016) provides three different methods for the determination of the ESC-OX impurities
in raw material and different dosage forms [12].
The main aims of this work were to develop simple, isocratic analytical method for the determination of ESC-OX
www.sscp-journal.com
© 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
1
BADULLA ET AL.
2
and five most common its impurities named as CIT A, B,
C, D, E and oxalic acid (OX-A) and to study the selectivity
of different columns with diverse characteristics as an effective alternative to the current official USP used and other
developed methods in order to achieve better resolution at a
short analysis time. Using columns of different chemical properties provides separation of unresolved peaks of two compounds [13]. Evaluation and comparison of the separation
performances of 12 columns were performed by using a system suitability test according to USP 2016. Six columns of
three different stationary phase chemistries (C18 (octadecyl),
pentafluorophenyl (PFP) and phenyl-hexyl) and two different
particle properties (i.e. core-shell and fully porous particle)
were selected for further comparative study. The validation
was carried out by using the three core-shell columns. The
selectivity of the six columns was evaluated by using correlation coefficients between the K values of all analytes [14].
The ESC-OX and its impurities are basic compounds. The
analysis of these compounds with conventional C18 column
requires acidic mobile phase, which may be problematic due
to the interaction of the residual silanol on the silica material.
These interactions may result in poor separation with peak
tailing which affect the resolution of the basic analytes. For
this reason, the separation power of stationary phases of different chemistry was examined. The effect of stationary phase
chemistry was evaluated by applying the same analytical and
instrumental parameters. The applicability of the established
method was evaluated by analyzing valid and expired dosage
forms.
2
2.1
M AT E R I A L S A N D M E T H O D S
Drugs and reagents
The ESC-OX and its impurities (Figure 1) were supplied
by the United States Pharmacopeia Convention, USA, with
99.2% purity. All mobile phase compositions were of HPLC
grade (Sigma-Aldrich, Switzerland).
2.2 Instrumentation and chromatographic
conditions
The method development was carried out in HPLC system
A, containing a gradient pump with the LC-20AT model
LPGE unit, SIL-20AC automatic sampler, CBM-20A model
communication unit, CTO-10-ASvp model column furnace,
SPD 20A model UV/VIS detector and DGU-20ASR model
degassing. An Intel Pentium 4 processor computer and Shimadzu LabSolutions, LC solution version 1.25 data analysis
program was used in addition to the HPLC product of Shimadzu (Japan). The pressure changed with column types, it
was between 258–285 bar for core-shell columns and between
142–180 bar for fully porous columns. The dead volume is
300 μL.
Other instruments used for ruggedness and specificity studies consist of the following parts: the gradient pump, SIL20A automatic sampler, CBM-20A model communication
unit, CTO-10-ASvp model column oven, SPD-M20A model
photo diode array detector (DAD) and DGU-20A5 model
The chemical structures of
escitalopram and its impurities
FIGURE 1
BADULLA ET AL.
8%
Reference: a (17), b (18), c (19), d (20), e (21).
11%
2–9
2–9
9%
Core-Shell particles
Core-Shell particles
200 m2 g−1
200 m g
1.5–8.5
1.5–8.5
100 Å
100 Å
2.6 μm
15 cm × 4.6 mm
15 cm × 4.6 mm
c
Kinetex PFP (phenomenex)
Kinetex® C18 (phenomenex)
2.6 μm
−1
2
c
®
Core-Shell particles
Fused-Core particle platform
150
200 m2 g−1
1.5–8.5
2–9
90 Å
100 Å
2.6 μm
2.7 μm
10 cm × 4.6 mm
15 cm × 4.6 mm
c
Express Phenyl-Hexyl (Supelco Analytical)
Kinetex® Phenyl-Hexyl (phenomenex)
g−1
m2
e
Ascentis®
19%
2–8
Fused-Core particle platform
Fused-Core particle platform
150 m2 g−1
150 m g
1–9
2–9
90 Å
90 Å
2.7 μm
10 cm × 4.6 mm
10 cm × 4.6 mm
e
Ascentis Express F5 (Supelco Analytical)
Ascentis® Express C18 (Supelco Analytical)
2.7 μm
−1
2
e
®
15%
Porous Silica
Fully Porous Silica
475 m2 g−1
1.5–7.5
80 Å
4 μm
15 cm × 4.6 mm
Synergi® Hydro-RP (Phenomenex)
350 m2 g−1
2–8
100 Å
3 μm
10 cm × 4.6 mm
d
Nucleosil® C18 (Teknokroma)
c
11.5
Ultra-pure, metal-free silica (99.99 % purity)
1.5–9.0
10%
11.5%
Ultra-pure, metal-free silica (99.99 % purity)
400 m g
400 m2 g−1
100 Å
100 Å
15 cm × 4.6 mm
3 μm
15 cm × 4.6 mm
c
Luna PFP (Phenomenex)
Luna® Phenyl-Hexyl (Phenomenex)
3 μm
1.5–9.0
−1
2
c
®
Matrix
High Purity, base deactivated
100 m2 g−1
450
2–7.5
2–8
100 Å
100 Å
3 μm
15 cm × 4.6 mm
5 μm
g−1
m2
Surface Area
pH Range
Pore Size
Particle Size
L x I. D.
10 cm × 4.6 mm
b
The comparison of the performances of 12 columns was performed by evaluation of the system suitability test according
to the USP 2016.
The specificity of the established method was determined
according to the ICH Q2(R1) guideline [20]. The chromatograms of the blank oral solution and the tablet matrix
were compared with the chromatograms of the cited analytes
by using three core-shell columns.
The linearity of the developed method was evaluated by
mean of three core-shell columns. The calibration curve for
ESC-OX and its impurities was carefully chosen in subsequent range ESC-OX (0.236–70.70 μg mL−1 ), CIT A
(0.0241–72.30 μg mL−1 ), CIT B (0.208–62.40 μg mL−1 ), CIT
C (0.202–60.60 μg mL−1 ), CIT D (0.208–62.40 μg mL−1 ) and
ODS-3 (GLSciences)
Method validation
Fluophase® RP (Thermo Scientific)
2.5
Physicochemical Properties of the Tested Columns
The mobile phase composition contains ACN/methanol
(MeOH)/H2 O/phosphate buffer (50 mM, pH = 3.5)
(25:5:20:50, v/v/v/v). The phosphate buffer solution (50 mM,
pH = 3.5) was prepared by dissolving 6.8 g of KHPO4 .2H2 O
and 115.6 μL of H3 PO4 (d = 1.695 g mL−1 and 86%)
in ultrapure water (V = 1000 mL). The pH of the buffer
solution was controlled by using a calibrated pH-meter.
After preparation, the solution was filtered through a 0.45
μm pore size cellulose membrane filter and sonicated for
15 min to remove dissolved gases. Mixing of the quantified
solvents at the anticipated percentages was carried out via the
low-pressure gradient mixing unit of the HPLC device.
a
Preparation of mobile phase
TABLE 1
2.4
Inertsil®
To prepare the standard stock solutions of the analytes, appropriate amount of standard was weighed and dissolved in
25 mL of ACN/water (30:70, v/v) solution. For this purpose,
3.01 mg of ESC-OX, 2.41 mg of CIT A, 2.08 mg of CIT
B, 2.02 mg of CIT C, 2.08 mg of CIT D, 2.06 mg of CIT
E and 2.57 mg of OX-A were used. Accordingly, the stock
solutions were mixed and diluted using the same diluent to
obtain the test solution used in the method development studies. The final concentrations of the analytes were as follows:
ESC-OX 15.05 μg mL−1 , CIT A 12.05 μg mL−1 , CIT B 10.40
μg mL−1 , CIT C 10.10 μg mL−1 , CIT D 10.40 μg mL−1 , CIT E
10.30 μg mL−1 and OX-A 12.85 μg mL−1 .
3 Series High Purity Silica Gel
Standards preparations
Column (Brand)
2.3
Carbon Load
degassing equipped with LP. The Intel Pentium 4 processor computer and the Shimadzu Lab Solution LC solution
v1.11 SP1 data analysis program were used in addition to
the HPLC product of Shimadzu (Japan). Twelve columns
were applied in this study as it is represented in the Table 1
[15–19].
15%
3
BADULLA ET AL.
4
The effect of mobile phase pH on the retention of ESC-OX and its impurities by using (column: Kinetex® PFP core-shell column
(3 μm, 150 × 4.6 mm), mobile phase; buffer: ACN: MeOH: H2 O (50:25:5:20, v/v/v/v). (a) pH = 2, (b) pH = 3, (c) pH = 3.5, all phosphate buffer,
(d) pH = 4(acetate Buffer), (e) pH = 5(acetate Buffer), (f) pH = 6 (phosphate Buffer)
FIGURE 2
CIT E (0.206–61.80 μg mL−1 ). The linearity was estimated by
linear regression analysis with intra- and interday replicates.
The LOD and LOQ values were calculated by using signal to
noise ratio. For each column, the (S/N) ratio attained from
the chromatograms is multiplied by 3 for the LOD and 10 for
the LOQ. Moreover, probable variations between the analytical columns were studied by one-way ANOVA test. Statistical calculations were completed with Graph Pad Prism v 6.0 b
software. The standard addition method was used to evaluate
the accuracy of the current method. Three different percentages (50, 100, 150%) of each analyte was added to the preanalyzed tablet and oral solution. All the concentration levels were prepared three times, spiked to the pharmaceutical
dosage forms and then the mean recovery was calculated.
The intraday and interday (intermediate) precision evaluations were performed at the lowest concentration by using
standard solutions for each analyte.
In order to estimate the method robustness. The changes in
mobile phase composition, concentration, buffer pH, wavelength, flow rate and temperature were followed according to
Vander Heyden [21]. The standard solutions of ESC-OX and
its impurities were examined for the short term (1 day), freeze
and thaw and long term (30 days).
3
RESULTS AND DISCUSSION
3.1 Selection of the maximum UV absorbance
wavelength
The scanning of the maximum absorbance at UV-region
was carried out between 200–400 nm for ESC-OX and its
impurities. All cited analytes had maximum absorbance in
196–210 nm region. Accordingly, 210 nm was selected for
further study.
3.2
Chromatographic conditions
The selection of mobile phase composition was performed
by using Phenomenex Kinetex® PFP core-shell column. All
tested analytes are basic compounds; the pKa value of ESCOX is 9.5 while the predicted pKa of its impurities is in the
following ranges 9.48–16.06. In order to get a reproducible
result, the ionization of the cited analytes should be controlled
by adding the buffer solution to the mobile phase. Several
buffer salts in the pH range of 2–6 were tested. The retention time of the ESC-OX and its impurities were extremely
affected by the mobile phase buffer pH. The examination of
pH effect was conducted by using mobile phase that contains,
BADULLA ET AL.
The chromatograms related to the method development by using Kinetex® PFP core-shell column (a) buffer: ACN: H2 O
(50:30:20) v/v/v/v, (b) buffer: ACN: H2 O (50:25:25, v/v/v/v), (c) buffer: ACN: MeOH: H2 O (50:30:5:15, v/v/v/v), (d) buffer: ACN: MeOH: H2 O
(50: 20:10:20, v/v/v/v), (e) buffer: ACN: MeOH: H2 O (50:25:5:20, v/v/v/v)
FIGURE 3
5
BADULLA ET AL.
6
TABLE 2
Results of system suitability tests (n = 3)
Kinetex® C18 Core-Shell Column.
Parameters CIT A
CIT B CIT C
CIT D
Inertsil ODS-3® C18 Fully Porous Column.
ESC-OX CIT E
CIT A
CIT B
CIT C
CIT D
ESC-OX CIT E
N
3189.13 803.68 12157.45 14817.70 14472.58 17438.25 3682.75 1006.07 9758.81 10339.27 10900.16 11265.25
tR (min)
2.105
3.404
4.882
5.753
6.194
7.228
2.992
5.443
8.265
9.802
10.614
12.865
T
1.263
1.001
1.234
1.215
1.254
1.169
1.237
1.011
1.362
1.380
1.416
1.417
As
1.255
1.004
1.217
1.202
1.239
1.160
1.230
1.012
1.326
1.343
1.370
1.364
K’
0.655
1.678
2.843
3.526
3.872
4.678
1.060
2.748
4.691
5.75
6.309
7.859
Rs
4.901
3.656
4.013
4.601
2.180
4.773
7.117
4.874
4.866
4.025
1.922
4.695
α
2.562
1.694
1.240
1.098
1.210
-
2.952
1.707
1.225
1.097
1.246
-
RSD% of tR
0.41
0.34
0.19
0.20
0.19
0.30
0.19
0.25
0.29
0.33
0.32
0.27
Kinetex® PFP Core-Shell Column.
Luna® PFP Fully Porous Column
Parameters
CIT A
CIT B
CIT C
CIT D
ESC-OX
CIT E
CIT A
CIT B
CIT C
CIT D
ESC-OX
CIT E
N
4415.28
769.69
15178.22
15987.17
14922.74
17851.06
5288.87
808.78
7407.11
7798.72
6884.89
6839.233
tR (min)
2.618
4.446
6.802
7.380
8.302
9.586
4.051
8.363
14.024
15.326
17.676
19.775
T
1.327
1.018
1.325
1.318
1.485
1.352
1.332
1.0389
1.359
1.357
1.416
1.398
As
1.297
1.023
1.293
1.290
1.447
1.314
1.294
1.034
1.320
1.319
1.374
1.360
K’
0.829
2.107
3.754
4.159
4.805
5.700
1.789
4.747
8.655
9.547
11.163
12.609
Rs
7.071
4.580
5.465
2.547
3.648
4.588
11.202
6.145
6.190
1.922
3.005
2.292
α
2.540
1.781
1.108
1.155
1.186
-
2.653
1.823
1.103
1.169
1.130
-
RSD% of tR
0.187
0.214
0.239
0.244
0.367
0.251
0.324
0.360
0.435
0.422
0.469
0.445
Kinetex® Phenyl-Hexyl Core-Shell Column.
Parameters CIT A
CIT B CIT C
CIT D
Luna® Phenyl-Hexyl Fully Porous Column.
ESC-OX CIT E
CIT A
CIT B
CIT C
CIT D
ESC-OX CIT E
N
3155.52 614.36 11998.06 14284.38 14990.14 15107.56 5878.93 834.011 15508.38 16635.45 16877.16 16249.16
tR (min)
2.272
3.697
5.242
6.072
6.712
7.861
3.399
6.353
9.585
11.307
12.632
14.620
T
1.335
1.001
1.277
1.252
1.336
1.470
1.237
1.023
1.108
1.109
1.149
1.151
As
1.310
1.003
1.257
1.241
1.315
1.394
1.214
1.010
1.105
1.105
1.147
1.140
K’
0.745
1.853
3.048
3.684
4.173
5.606
1.386
3.461
5.731
6.941
7.871
9.266
Rs
5.695
3.758
3.920
4.207
3.027
4.836
10.301
5.588
5.441
5.232
3.581
4.689
α
2.487
1.043
1.208
1.133
1.343
-
2.497
1.656
1.211
1.134
1.177
-
RSD% of tR
0.067
0.041
0.029
0.025
0.031
0.048
0.158
0.286
0.354
0.336
0.355
0.348
buffer: ACN: MeOH: H2 O (50:25:5:20, v/v/v/v). Since ESCOX and its impurities are basic compounds, these compounds
are relatively ionized by using a buffer solution of pH 2–6
and so their retention times were reduced. By using phosphate
buffer of pH = 2 the shape of CIT B was odd. This can be
explained by interconversion between chiral isomers of CIT
B [22]. By changing the buffer pH to 3 the peak shape of
CIT B became broad. Using a buffer solution of pH 4, the
peak of CIT B was sharp; however, the ESC-OX and CIT
E eluted together. The retention times of the analytes in the
buffer pH range from 2–4 depended mainly on their polarities
(log p). At pH 5 and pH 6, CIT C and CIT E co-eluted. The
elution order was changed at pH 5 and 6, the CIT E eluted
between CIT C and CIT D, ESC-OX was the latest eluting
compound in contrast to the pH 2 to 3.5, where CIT E was
the latest analyte. Alteration of the sequence of elution can
be clarified by variation in the ionization degree of the compounds and electrostatic interaction between the F atom in
PFP stationary phase and amine lone pair in ESC-OX and
its impurities [23]. Using buffer of higher pH leads to delay
in the retention time of all analytes. This can be clarified by
the decrease in the degree of ionization, and may be due to
slight π-π interaction between the aromatic rings of analytes
and stationary phase [24] and some electrostatic interaction
or by strong adsorption of compounds with the stationary
phase [25]. However, the particular mechanism for the multipart interaction of these compounds with stationary phase
by changing pH is complex and yet unsolved [26]. It is clear
from the pH scanning that the retention times of all analytes
and the peak shapes are strongly influenced by the pH value
as shown in Figure 2. By using potassium phosphate buffer of
pH 3.5 (50 mM), the peak shape of CIT B was sharper with a
good resolution and reasonable retention time. Therefore, further analyses were carried out at this pH. Various percentages
of ACN were tested and the best result was obtained at 25%
by volume, but the resolution between the CIT C and CIT D
BADULLA ET AL.
7
Chromatogram of standard solution of ESC-OX and its impurities by using (a) Kinetex® C18 (2.6 μm, 150 × 4.6 mm) core-shell
column, (b) Inertsil® C18 (3 μm, 150 × 4.6 mm) fully porous column, (c) Kinetex® PFP (2.6 μm, 150 × 4.6 mm) core-shell column, (d) Luna® PFP
(2.6 μm, 150 × 4.6 mm) fully porous column, (e) Kinetex® phenyl-hexyl (2.6 μm, 150 × 4.6 mm) core-shell column, (f) Luna® phenyl-hexyl
(2.6 μm, 150 × 4.6 mm) fully porous column
FIGURE 4
was low; thus, to improve the resolution between these compounds and peak symmetry, and reduce tailing of CIT E peak,
MeOH was added. MeOH has an UV cut-off of 205 nm. For
getting absorbance of MeOH less than 0.05 AU (absorbance
unit), the working wavelength must be chosen to be greater
than 235 nm or the MeOH percentage must be lower than
15%. The best result was attained by using the buffer of ACN:
MeOH: H2 O (50:25:5:20, v/v/v/v). In addition, various gradient elution was also tried, however, the best result obtained by
isocratic elution which is preferred over the gradient elution
due to not requiring special instrument, highly trained technician and easy to prepare.
The effect of flow rate is relatively small on the resolution, but it affects the overall analysis time and last peak’s
tailing. The flow rates of 1.1, 1.2 and 1.3 mLmin−1 were
tried. The reasonable analysis time with the best peak shape
was acquired for the 1.2 mLmin−1 flow rate. The temperature
should be kept constant during analysis time because it affects
selectivity. The column oven temperature was varied as 35, 40
and 45◦ C. The column oven temperature kept at 40◦ C because
no valuable decrease in run time was observed at the higher
temperature. This temperature was best for maintaining low
viscosity of the mobile phase and back pressure, providing
fast flow rate and lower the analysis time. In addition, two
buffer concentrations were tried (25 and 50 mM). 50 mM was
selected because the peaks were sharper, especially for CIT B.
The chromatograms related to the method development are
given in Figure 3.
3.3
Selection of the columns
Selection of the columns plays important role in getting a
powerful separation with reasonable resolution. In this work,
12 columns with different length and particle properties were
8
examined (fully porous and core-shell). After evaluation of
the separation power of all columns by using SST (Table 2),
six best performing columns were selected for further
study.
The choice of PFP column was established on the existence
of fluorine (F) in the stationary phase to enhance the retention and selectivity of the halogenated polar compounds [19].
In addition, this stationary phase acts as Lewis acid (electron acceptor) and offers π-π interaction, dipole-dipole interaction, steric and shape/size selectivity [27]. Choosing the
phenyl-hexyl stationary phase was established on having variant selectivity for aromatic and hydrophobic interactions. This
selectivity comes from the existence of six-carbon chain and
aromatic phenyl group. Phenyl is considered as a Lewis base
(electron donor) [19], a good source of selectivity for the
amine and heterocyclic containing aromatic compound and
π-π interaction [28]. Two different chemistry stationary
phases (PFP and phenyl-hexyl) were selected and compared
with the traditional C18 column. The idea of using different
BADULLA ET AL.
physical and chemical stationary phases was found useful to
evaluate the selectivity of columns.
The elution order of all compounds was the same in all of
the columns; only a trivial superiority of phenyl hexyl stationary phase was detected in term of resolution. This may
be originated from different multisite interactions of the compounds with the phenyl ring. Relatively longer retention in the
PFP stationary phase in comparison with the other two stationary phases is due to the presence of the F atom which can
form dipole-dipole interactions and H-bonds with the solutes,
resulting in slightly longer retention of the solutes. Also, the
longer retention may be due to the larger size of fluorinatedphenyl ring in comparison to the C18 and phenyl-hexyl ligands of the stationary phase [29]. This leads to increase steric
retention of the solutes. The peak symmetry was relatively
better by using phenyl-hexyl stationary phase, especially for
CIT B. The chromatograms relating to the 12 columns are
shown in Figure 4 and 5. The result of comparing the system
suitability test of the six-selected columns (Table 2) indicated
Chromatogram of ESC-OX and its impurities by using;(a) Nucleosil® 100 C18 (3 μm, 10 × 0.46 cm), (b) Synergi Hydro-RP®
(4 μm, 15 × 0.46 cm), (c) Fluophase RP® (3 μm, 10 cm × 4.6 mm), (d) Ascentis® Express C18 (2.7 μm, 10 cm × 4.6 mm), (e) Ascentis® Express F5
(2.7 μm, 10 cm × 4.6 mm), (f) Ascentis® Express Phenyl-hexyl (2.7 μm, 10 cm × 4.6 mm)
FIGURE 5
BADULLA ET AL.
that all core-shell columns provide excellent separation of the
ESC-OX and its impurities in regard to analysis time, theoretical plate and resolution.
The core-shell columns have the highest number of theoretical plates with the preference of PFP column, the
retention times were almost similar in the three core-shell
columns. Comparing, the capacity factor (K’) and selectivity
of columns (α) the best results obtained by using fully porous
columns that provide suitable analysis time with good selectivity but the peaks were slightly broader. The difference in
the tailing factor of all columns was not great and all values
were within the recommended value (T < 2). The resolution
values between the CIT D, ESC-OX and CIT E were close
to each other in most columns with the slight superiority of
phenyl-hexyl column. In concern to the entire columns’ performance, all core-shell columns provide excellent separation
of the ESC-OX and its impurities in regard to analysis time,
theoretical plate and resolution.
The dissimilarity factor of the core-shell column with the
fully porous column was estimated by mean of regression
analysis and correlation coefficients (r) between the K’ values
of all analytes. The low correlation coefficient shows the high
dissimilarity and vice versa [30]. A plot of K’ achieved from
Inertsil® C18 fully porous column versus the K’ achieved
from Kinetex® C18 core-shell column refers a high degree
of correlation (r = 0.99959), i.e. high degree of similarity as
9
shown in Figure 6 (a). By relating the correlation between K’
achieved from Luna® PFP fully porous column versus the K’
achieved from Kinetex® PFP core-shell column and Luna®
phenyl-hexyl fully porous column versus the K’ achieved from
Kinetex® phenylhexyl core-shell column the result displays
that the values are, r = 99872 and r = 0.99110, respectively
(Figure 6 b, c). Consequently, it can be inferred that the similarity in the selectivity between core-shell column and fully
porous column for ESC-OX and its impurities order as follows: C18 > PFP > phenyl-hexyl.
The selectivity of stationary phase chemistry of conventional C18 with the PFP and phenyl-hexyl stationary phases
were compared by the similar manner and the correlation
coefficient between the C18 with the PFP and C18 with
phenyl-hexyl stationary phases indicate that the similarity
of PFP column selectivity is close to the C18 more than
the phenyl-hexyl column. The correlation coefficients are,
r = 0.9970 and r = 0.9921 (Figure 6 (d)). The result of the
selectivity approves the slight superiority of phenyl-hexyl column in the separation for the ESC-OX and its impurities over
the conventional C18 and PFP. However, there is an insignificant dissimilarity between the C18 traditional column and the
other two stationary phases which refers to the ability of using
traditional C18 in separation of ESC-OX and its impurities in
case of unavailability of other columns with specific chemical
ligands.
k’ of ESC-OX and its impurities on (a) C18 fully porous column versus C18 core-shell column, (b) PFP fully porous column
versus PFP core-shell column, (c) phenyl-hexyl fully porous column versus phenyl-hexyl core- shell column, (d) K’ of ESC-OX and its impurities on
C18 column versus PFP and phenyl-hexyl core-shell column versus PFP Column
FIGURE 6
10
TABLE 3
Statistical Evaluation of Linearity and precision studies
Kinetex® Phenyl-Hexyl Core-Shell Column.
Linearity range (μg
mL−1 )
Slope ± SDa (intraday, n = 27)
CIT A
CIT B
CIT C
CIT D
ESC-OX
CIT E
0.0241–72.30
0.208–62.40
0.202–60.60
0.208–62.40
0.236–70.70
0.206–61.80
36667 ±52.60
24005 ± 165.8
63780 ± 191.50
30544 ± 43.41
37942 ± 72.95
27851 ± 41.07
Intercept ± SDa (intraday, n = 27)
11214 ± 1598
3304 ± 4346
4445 ± 4876
2547 ± 1138
1672 ± 2167
-5770 ± 1066
Regression coefficient (intraday, n = 27)
0.9999
0.9988
0.9998
0.9999
0.9999
0.9999
95% CI of Slope
36558 to 36775
23664 to 24347
63385 to 64174
30455 to 30634
37792 to 38092
27767 to 27936
95% CI of Intercept
7922 to 14506
−5649 to 12257
−5598 to 14489
201.9 to 4892
−2792 to 6137
−7967 to −3573
LOQ
0.005
0.008
0.003
0.006
0.006
0.007
LOD
0.016
0.025
0.009
0.019
0.017
0.021
Repeatability (inter-day, mean, n = 18)
13835.56
4344.83
10302.94
6322.72
7249.17
3631.72
Repeatability (inter-day, RSDc %, n = 18)
0.71
0.86
0.62
0.87
0.76
1.15
Repeatability (inter-day, SDa, n = 18)
97.95
37.13
63.86
55.01
55.16
41.81
Repeatability (intraday, SEMd, n = 18)
39.99
15.16
26.07
22.46
22.52
17.07
Repeatability (inter-day, CIe, n = 18)
13732.79 to 13938.32
4301.93 to 4301.93
10235.95 to 10369.94
6265.01 to 6380.43
7191.30 to 7307.04
3587.85 to 3675.59
Kinetex® PFP Core-Shell Column.
CIT A
CIT B
CIT C
CIT D
ESC-OX
CIT E
Linearity range (μg mL−1 )
0.0241–72.30
0.208–62.40
0.202–60.60
0.208–62.40
0.236–70.70
0.206–61.80
Slope ± SDa (intraday, n = 27)
40898 ± 54.94
25068 ± 93.82
64661 ± 96.28
30168 ± 51.71
38535 ± 35.59
27162 ± 39.32
Intercept ± SDa (intraday, n = 27)
1874 ± 1669
−467.8 ± 2088
−5340 ± 2452
207.7 ± 1356
−454.2 ± 1057
2158 ± 1021
1.000
0.9996
0.9999
0.9999
1.000
0.9999
95% CI of Slope
−1564 to 5312
24874 to 25261
64463 to 64860
30062 to 30275
38462 to 38609
27081 to 27243
95% CI of Intercept
−0.1301 to 0.0382
−4769 to 3833
−10391 to -289.6
−2586 to 3001
−2632 to 1724
54.51 to 4262
LOQ
0.013
0.020
0.008
0.017
0.017
0.018
LOD
0.038
0.061
0.024
0.050
0.050
0.055
Repeatability (inter-day, mean, n = 18)
11611.50
5618.56
11114.39
5654.22
9349.67
6683.22
Repeatability (inter-day, RSDc %, n = 18)
1.22
1.77
0.63
1.26
1.19
1.31
Repeatability (inter-day, SDa, n = 18)
141.92
99.37
70.14
71.38
111.71
87.51
Repeatability (intraday, SEMd, n = 18)
57.94
40.57
28.63
29.14
45.61
35.72
Repeatability (inter-day, CIe, n = 18)
11497.94 to 11760.40
5514.30 to 5722.81
11040.80 to 11187.98
5579.33 to 5729.12
9232.46 to 9466.88
6591.41 to 6775.03
(Continues)
BADULLA ET AL.
Regression coefficient (intraday, n = 27)
BADULLA ET AL.
TABLE 3
(Continued)
Kinetex® C18 Core-Shell Column.
CIT A
CIT B
CIT C
CIT D
ESC-OX
CIT E
Linearity range (μg mL−1 )
0.0241–72.30
0.208–62.40
0.202–60.60
0.208–62.40
0.236–70.70
0.206–61.80
Slope ± SDa (intraday, n = 27)
41326 ± 32.75
23086 ± 91.56
67498 ± 101.80
29412 ± 201.0
37468 ± 54.35
24007 ± 75.61
Intercept ± SDa (intraday, n = 27)
−4896 ± 995.1
1435 ± 2401
−1067 ± 2592
3013 ± 5271
35.50 ± 1615
274.4 ± 1964
Regression coefficient (intraday, n = 27)
1.000
0.9996
0.9999
0.9988
0.9999
0.9998
95% CI of Slope
41258 to 41393
22897 to 23274
67288 to 67708
28998 to 29826
37356 to 37580
23851 to 24163
95% CI of Intercept
−6946 to -2847
−3511 to 6381
−6406 to 4272
−7844 to 13871
−3290 to 3361
−3771 to 4319
LOQ
0.004
0.008
0.003
0.006
0.006
0.007
LOD
0.014
0.023
0.008
0.018
0.018
0.021
Repeatability (inter-day, mean, n = 18)
10309.40
5437.06
12678.17
6207.33
8710.16
6175.66
Repeatability (inter-day, RSDc %,
n = 18)
0.79
1.73
0.73
1.18
1.21
0.98
Repeatability (inter-day, SDa, n = 18)
80.99
94.02
92.66
73.08
105.02
60.81
Repeatability (intraday, SEMd, n = 18)
33.06
38.38
37.83
29.83
42.87
24.82
Repeatability (inter-day, CIe, n = 18)
10224.43 to
10394.37
5338.41 to
5535.70
12580.95 to
12775.38
6130.66 to
6284.01
8599.98 to
8820.35
6058.20 to
6185.80
ANOVA (between columns)
F (1.004,
8.035) = 3.276,
P = 0.1077
F (1.011,
8.084) = 3.818,
P = 0.0858
F (1.010,
8.084) = 4.238,
P = 0.0729
F (1.019,
8.153) = 3.683,
P = 0.0902
F (1.173,
9.384) = 4.026,
P = 0.0702
F (1,010,
8.080) = 4.166,
P = 0.0750
11
BADULLA ET AL.
12
TABLE 4
Results of Accuracy Studies in Tablet Matrix and Oral Solution by Using Kinetex® Phenyl-Hexyl Core-Shell Column
Kinetex® Phenyl-Hexyl Core-Shell Column.
Tablet Matrix
Compounds
CIT A
CIT B
CIT C
CIT D
ESC-OX
CIT E
TABLE 5
Added
(𝛍g mL−1 )
Oral solution
Found
(𝛍g mL−1 )
Recovery %
Average
Recovery
98.87
2.41
2.36
98.04
4.82
4.80
99.53
7.23
7.16
99.04
2.08
2.14
103.04
103.44
Added
(𝛍g mL−1 )
Found
(𝛍g mL−1 )
Recovery %
Average
Recovery
97.92
2.41
2.40
99.53
4.82
4.66
96.65
7.23
7.055
97.58
2.08
2.11
101.24
4.16
4.34
104.23
4.16
4.34
104.22
6.24
6.43
103.06
6.24
6.48
103.84
2.02
2.02
99.80
4.04
4.08
101.04
6.06
6.05
99.92
2.08
2.04
98.29
100.2
100.77
2.02
1.96
97.03
4.04
4.06
100.48
6.06
6.03
99.50
2.08
2.04
97.86
4.16
4.32
103.86
4.16
4.11
98.84
6.24
6.25
100.18
6.24
6.25
100.12
2.35
2.33
98.73
4.71
4.83
102.41
7.07
7.03
99.39
2.06
2.09
101.45
100.17
101.08
2.35
2.39
101.36
4.71
4.82
102.38
7.07
7.45
105.36
2.06
2.08
100.97
4.12
4.22
102.42
4.12
4.15
100.62
6.18
6.14
99.39
6.18
6.28
101.62
103.10
99.00
98.94
101.62
101.07
Results of Accuracy Studies in Tablet Matrix and Oral Solution by Using Kinetex® PFP Core-Shell Column
Kinetex® PFP Core-Shell Column
Tablet Matrix
Compounds
CIT A
CIT B
CIT C
CIT D
ESC-OX
CIT E
Added
(𝛍g mL−1 )
Oral solution
Found
(𝛍g mL−1 )
Recovery %
Average
Recovery
99.99
2.41
2.44
101.14
4.82
4.84
100.37
7.23
7.12
98.45
2.08
2.12
101.88
101.10
Added
(𝛍g mL−1 )
Found
(𝛍g mL−1 )
Recovery %
Average
Recovery
99.86
2.41
2.46
101.92
4.82
4.77
99.02
7.23
7.13
98.63
2.08
2.08
100.22
4.16
4.15
99.73
4.16
4.03
100.91
6.24
6.33
101.36
6.24
62.8
101.24
2.02
2.06
101.74
4.04
3.97
98.21
6.06
6.02
99.35
2.08
2.10
100.56
99.77
99.54
2.02
2.04
101.16
4.04
4.02
99.49
6.06
6.01
99.17
2.08
2.12
101.74
4.16
4.12
99.15
4.16
4.13
99.22
6.24
6.17
98.90
6.24
6.24
99.93
2.35
2.34
99.50
4.71
4.86
103.11
7.07
7.20
101.86
2.06
2.11
102.25
101.49
101.15
2.35
2.43
103.14
4.71
4.85
102.96
7.07
7.25
102.65
2.06
2.11
102.25
4.12
4.21
102.11
4.12
4.21
102.11
6.18
6.19
100.18
6.18
6.23
100.77
100.79
99.94
100.30
102.92
101.71
BADULLA ET AL.
TABLE 6
13
Results of Accuracy Studies in Tablet Matrix and Oral Solution by Using Kinetex® C18 Core-Shell Column
Kinetex® C18 Core-Shell Column
Tablet Matrix
Compounds
CIT A
CIT B
CIT C
CIT D
ESC-OX
CIT E
3.4
Added
(𝛍g mL−1 )
Oral solution
Found
(𝛍g mL−1 )
Recovery %
Average
Recovery
104.17
2.41
2.54
105.27
4.82
4.96
103.00
7.23
7.54
104.25
2.08
2.21
106.47
104.85
Added
(𝛍g mL−1 )
Found
(𝛍g mL−1 )
Recovery %
Average
Recovery
105.58
2.41
2.59
106.58
4.82
5.02
104.21
7.23
7.66
105.96
2.08
2.16
103.82
4.16
4.35
104.45
4.16
4.33
104.18
6.24
6.47
103.62
6.24
6.68
107.00
2.02
1.99
98.29
4.04
3.90
96.58
6.06
5.84
96.30
2.08
2.15
103.41
97.06
102.41
2.02
2.04
100.83
4.04
3.92
96.98
6.06
5.87
96.93
2.08
2.08
99.76
4.16
4.21
101.27
4.16
4.12
99.08
6.24
6.40
102.56
6.24
6.31
101.15
2.35
2.38
100.84
4.71
4.59
97.48
7.07
6.92
97.84
2.06
2.11
102.56
98.72
103.06
2.35
2.37
100.45
4.71
4.63
98.33
7.07
7.05
99.78
2.06
2.18
105.89
4.12
4.28
103.79
4.12
4.20
101.83
6.18
6.35
102.82
6.18
6.37
103.00
Method validation
The result of the specificity test showed that there were no
interfering peaks at the retention time of the examined compounds. As a complement to the specificity tests, stability
studies were carried out. Short term (1 day), freeze and thaw
and long term (30 days) stability studies and the monitoring
of the stability showed that all compounds were stable at all
studied conditions.
The linearity was estimated by linear regression analysis
by using nine concentrations for all analytes by using the
three core-shell columns. The statistical data on the intraday and inter-day linearity of the method for the three coreshell columns are given in Table 3. In addition, possible
differentiations between the analytical columns were examined by one-way ANOVA test. The results revealed that there
were no noteworthy statistical differences among the analytical columns for all analytes.
The precision of the methods was studied at two steps:
intraday (repeatability) and interday (intermediate) precision.
The results showed that the method was analytically adequate
(RSD% < 2.0) in relation to the precision. The precision data
are outlined in Table 3. The developed method accuracy was
evaluated by the standard addition method for both dosage
form matrixes and the results were revealed that the current
method is sufficiently accurate (Table 4–6).
105
98.25
100.00
99.52
103.57
The robustness and raggedness evaluation outcome
revealed that the developed method was sufficiently robust
with RSD% below 2%. Moreover, the ruggedness of the
method was evaluated by the other HPLC system. There was
no significant difference between the chromatograms of the
two HPLC systems.
The applicability of the developed method was assessed
by using it for the determination of ESC-OX and its impurities (if any) in different pharmaceutical dosage forms (tablet
and oral solution). The statistical result of different dosage
forms analyses by three core-shell columns are summarized
in Table 7. At the same time, an expired oral solution (expiration date: April 2015) and tablet dosage form (expiration
date: August 2016) were examined to determine the presence
of impurities. The results obtained from different pharmaceutical dosage forms indicated the presence of CIT C impurity
in all dosage forms with a very low quantity. The impurity
CIT C is considered as one of the critical impurities that can
arise during cyclization step and it might remain after conventional purification steps [31]. In the tablet dosage form, CIT C
and CIT A were detected. Impurity CIT A may be resulted
from hydrolysis [32] and also monitored during synthesis of
the ESC-OX [33].
In the evaluation of the expired oral solution, dosage form
impurity CIT D was detected alongside impurity CIT C. This
CIT D was detected and identified as a degradation product of
BADULLA ET AL.
14
TABLE 7
Assay Results of ESC-OX pharmaceutical dosage forms
ESC-OX 20 mg Film Tablet (n = 6)
C18 column
PFP column
phenyl-hexyl column
Average (mg)
20.06
20.07
20.12
Minimum (mg)
20.00
20.00
20.07
Maximum (mg)
20.10
20.10
20.19
SD (mg)
0.032
0.034
0.046
RSD %
0.16%
0.17%
0.23%
SEM (mg)
0.0104
0.0107
0.0189
Bias%
−0.30
−0.35
−0.60
95% CI of Average
20.03 to 20.08
20.04 to 20.09
20.07 to 20.17
% Content Uniformity
100.3
100.35
100.6
ANOVA
F (2, 27) = 3.066, P = 0.0631
C18 column
PFP column
phenyl-hexyl column
10.34
10.30
10.29
ESC-OX 10 mg Film Tablet (n = 6)
Average (mg)
Minimum (mg)
10.27
10.16
10.20
Maximum (mg)
10.39
10.42
10.33
SD (mg)
0.042
0.072
0.037
RSD %
0.41%
0.70%
0.36%
SEM (mg)
0.0133
0.0228
0.0117
Bias%
−3.4
−3.0
−2.9
95% CI of Average
10.31 to 10.37
10.25 to 10.35
10.27 to 10.32
% Content Uniformity
100.3
100.35
100.2
ANOVA
F (2, 27) = 2.009, P = 0.1537
ESC-OX Oral Solution (10 mg mL−1 )
C18 column
PFP column
Average (mg)
10.23
10.21
phenyl-hexyl column
10.25
Minimum (mg)
10.18
10.14
10.21
Maximum (mg)
10.33
10.27
10.29
SD (mg)
0.054
0.043
0.025
RSD %
0.53%
0.42%
0.24%
SEM (mg)
0.0170
0.0135
0.0079
Bias%
−2.3
−2.1
−2.5
95% CI of Average
10.19 to 10.27
10.18 to 10.24
10.23 to 10.27
% Content Uniformity
102.3
102.1
102.5
ANOVA
CIT (racemic mixture) [34]. The examination of the expired
tablet, identified only impurity CIT A (Figure 7). The detected
quantities of all impurities were below the maximum allowed
limit of USP 2016.
4
CONC LU SI ON
In this study, a simple isocratic, reversed phase HPLC method
with a run time of 12 min was developed for quantitative
F (2,27) = 1.835,
P = 0.1789
determination of ESC-OX and its impurities. Moreover, the
method can be successfully applied for detection of impurities
in the pharmaceutical dosage forms. The developed method
was applied by using 12 different columns. Thereafter, six different columns (core-shell and fully porous) were examined
with three different stationary phases. The best separation was
obtained from core-shell columns with good peak symmetry,
resolution and short analysis time. The phenyl hexyl stationary phase with middle hydrophobic selectivity in comparison with the C18 and PFP stationary phase provided the best
BADULLA ET AL.
15
Chromatograms represent impurities in (a) an expired oral solution, (b) oral solution by using phenyl-hexyl column, (c) an expired
tablet dosage form 20 mg, (d) tablet dosage form 20 mg by using phenyl-hexyl column
FIGURE 7
resolution. The PFP fully porous stationary phase proved to
be more retentive with poor resolution and peak shape; however, the developed method can be utilized by using C18 fully
porous column if they are the only available choice with relatively longer run time about 15 min.
The method was fully validated and the results obviously show that it is accurate and precise with low LOD,
which makes it a preferable alternative from the official USP
method. By the development of this method, both objectives
of this study were achieved with satisfactory results.
ACKNOW LEDGMENTS
The authors acknowledge the financial support provided by
the Anadolu University project No: 1502S091, Eskişehir,
Turkey.
CO NFLICT OF I N T E R E ST
The authors have declared no conflict of interest.
ORC ID
Hassan Y. Aboul-Enein
https://orcid.org/0000-0003-0249-7009
REFERENCES
1. Brown, W. A., Khan, A., Which depressed patients should receive
antidepressants? CNS Drugs. 1994, 1, 341–347. https://doi.org/
10.2165/00023210-199401050-00004.
2. Görög, S., Laukó, A., Herényi, B., Estimation of impurity profiles
in drugs and related materials. J. Pharm. Biomed. Anal. 1988, 6,
697–705. https://doi.org/10.1016/0731-7085(88)80081-5
3. Rao, R. N., Raju, A. N., Nagaraju, D., Development and validation of a liquid chromatographic method for determination of enantiomeric purity of citalopram in bulk drugs and
pharmaceuticals. J. Pharm. Biomed. Anal. 2006, 41, 280–285.
https://doi.org/10.1016/j.jpba.2005.10.025
4. Sungthong, B., Jac, P., Scriba, G. K. E., Development and validation of a capillary electrophoresis method for the simultaneous determination of impurities of escitalopram including
the R-enantiomer. J. Pharm. Biomed. Anal. 2008, 46, 959–965.
https://doi.org/10.1016/j.jpba.2007.05.029
5. Dhaneshwar, S. R., Mahadik, M. V., Column liquid
chromatography-ultraviolet and column liquid chromatography/mass spectrometry evaluation of stress degradation behavior
of escitalopram oxalate. J. AOAC Int. 2009, 92, 138–147.
6. Ramana, B., Sharmaa, B. A., Ghugare, P. D., Nandavadekar, S.,
Singh, D., Karmuse, P. K., Kumar, A., Structural elucidation of process-related impurities in escitalopram by LC/ESIMS and NMR. J. Pharm. Biomed. Anal. 2010, 53, 895–901.
https://doi.org/10.1016/j.jpba.2010.06.019
7. Kaleemullah, T., Ahmed, M., Sharma, H. K., Rajput, P., Reverse
phase liquid chromatographic method for the quantification of dip-toluoyl-d-tartaric acid in escitalopram oxalate drug substance.
Eurasian J. Anal. Chem. 2011, 6, 197–205.
8. Soliman, S. M., Youssef, N. F., Enantiomeric Thin-Layer Chromatographic Assay of Escitalopram in Presence of “In-Process
Impurities”. J. Plan Chromatogr. 2011, 24, 475–481.
9. Dighe, V. V., Pawaskar, P., Adhyapak, S., Shambhu, N.,
Mestry, D., Development of normal phase chiral liquid chromatographic method for estimation of escitalopram oxalate and
determination of R-citalopram enantiomer from escitalopram
BADULLA ET AL.
16
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
oxalate in bulk drug and tablet. J. Chem. Pharm. Res. 2012, 4,
4804–4809.
Deng, X., De Wolf, J., Vervoort, R., Pamperin, D., Adams, E.,
Schepdae, A. V., Development and validation of a capillary electrophoresis method for the determination of escitalopram and sensitive quantification of its enantiomeric impurity in formulations.
Electrophoresis. 2012, 33, 1648–1651.
Vaghela, B. K., Rao, S. S., Development and validation of stability
indicating RP-LC, short runtime method for the estimation of escitalopram in escitalopram dosage form. World J. Pharm. Res. 2013,
2, 1018–1030.
U.S. Pharmacopoeia-National Formulary (USP 39 NF 34), Volume
2. Escitalopram Oxalte monograph. (p. 3765–3770). USA: Book
Press, Inc., Baltimore, MD 2016, http://www.usp.org/.
Ng, L., Lunn, G., Faustino, P., in: R. J., Smith and M. L., Webb
(Eds.), Organic impurities in drug substance: origin, control, and
measurement in Analysis of Drug Impurities. Blackwell Publishing,
UK 2007, pp. 12.
Van Gyseghem, E., Van Hemelryck, S., Daszykowski, M.,
Questier, F., Massart, D. L., Vander Heyden, Y., Determining
orthogonal chromatographic systems prior to the development of
methods to characterize impurities in drug substances. J. Chromatogr. A. 2003, 988, 77–93.
Inertsil ODS-3. http://www.glsciences.com/c-product/hplc/lccolumns/inertsil-ods-3. (accessed, December 8, 2016).
Lupo, S., Carroll, F., Method development and column Selection:
How the fluorophenyl phase provides the power of HILIC and
reversed-phase modes in One Column. https://www.restek.com/
pdfs/GNAR2613-UNV.pdf (accessed, Nonember18, 2016).
Gritti, F., Guiochon, G., Kinetic Core shell column. LCGC
(North America) 2012, 30, 586–595. https://www.phenomenex.
com/Kinetex/CoreShellTechnology. (accessed, December 2,
2016).
Nucleosil standard C18 Phases. https://www.mnnet.com/HPLC
Start/NUCLEOSILphases/NUCLEOSILC18/tabid/6128/language/
en-US/Default.aspx (accessed, October 9, 2016).
https://www.drugs.com/sfx/escitalopram-side-effects.html. (accessed, December 23, 2016).
ICH-Topic Q2 (R1), Validation of Analytical Procedures: Text and
Methodology. Harmonized Tripartite Guideline, 2005, pp. 1–13.
http://www.ich.org/pdfICH/Q2A.
Vander Heyden, Y., Nijhuis, A., Smeyers-Verbeke, J., Vandeginste, B. G. M., Massart, D. L., Guidance for robustness: ruggedness tests in method validation. J. Pharm. Biomed. Anal. 2001, 24,
723–753.
Kern, W. J., Kirkland, K., in: R. L. Snyder, J. Kirkland, J.
L. Giajch (Eds.). Chiral separations in Practical HPLC method
development. A Wiley-Interscience publication, New York 1997,
pp. 562.
Marin, A., Barbas, C., Systematic comparison of different functionality columns for a classical pharmaceutical problem. J. Pharm.
Biomed. Anal. 2006, 40, 262–270.
24. Przybyciel, M., Santangelo, M. A., Paper Presented at the 54th
Pittsburg Conference on Analytical Chemistry and Applied Spectroscopy. Orlanda, Florida, USA 2003, 914, 210–215.
25. Sun, M., Liu, D. Q., in: P. G. Wang and W. He (Eds.), Analysis
of pharmaceutical impurities using hydrophilic interaction liquid
chromatography in Hydrophilic Interaction Liquid Chromatography
(HILIC) and Advanced Applications. CRC Press. Tayler and Francis Group, USA 2011, pp. 279.
26. Przybyciel, M., Novel Phases for HPLC Separations. Recent Developments in LC Column Technology. 2003, 2–5.
27. Yamamoto, F. M., Rokushika, S., Retention properties of the fluorinated bonded phase on liquid chromatography of aromatic hydrocarbons. J. Chromatogr. A. 2000, 898, 141–151.
28. Goss, J. D., Improved liquid chromatography of salicylic acid and
some related compounds on a phenyl column. J. Chromatogr. A.
1998, 828, 267–271.
29. Petruczynik, A., Wróblewski, K., Dzioba, K., WaksmundzkaHajnos, M., Retention, separation selectivity and system efficiency
of selected basic psychotropic drugs on different RPLC columns.
Open Chem. 2015, 13, 943–950.
30. Wilson, N. S., Gilroy, J., Dolan, J. W., Snyder, L. R., Column selectivity in reversed-phase liquid chromatography VI. Columns with
embedded or end-capping polar groups. J. Chromatogr. A. 2004,
1026, 91–100.
31. Vedantham, R., Vetukuri, V. P. R., Boini, A., Khagga, M.,
Bandichhor, R., Improved one-pot synthesis of citalopram diol and
its conversion to citalopram. Org. Process Res. Dev. 2013, 17,
798–805.
32. Sharma, M., Jawa, P. R., Gill, R. S., Bansal, G., Citalopram
hydrobromide: degradation product characterization and a validated
stability-indicating LC-UV method. J. Braz. Chem. Soc. 2011, 22,
836–848.
33. Gangula, S., Kolla, N. K., Elati, C. R., Dongamanti, A.,
Bandichhor, R., Identification, synthesis and spectral characterization of impurities in process development of Escitalopram. GERF
Bulletin of Biosciences. 2011, 2, 1–5.
34. Tadic, S., Nikolić, K., Agbaba, D., Development and optimization
of an HPLC analysis of citalopram and its four nonchiral impurities using experimental design methodology. J. AOAC Int. 2012, 95,
733–743.
How to cite this article: Badulla WFS, Can
NÖ, Atkosar Z, Arli G, Aboul-Enein HY. Comparative Study of Different Chemistries and Particle
Properties, High-Performance Liquid Chromatography Stationary Phases in Separation of Escitalopram Oxalate and Its Impurities in Different Pharmaceutical Dosage Forms. Sep Sci plus. 2019;1-16.
https://doi.org/10.1002/sscp.201900021